Modified Variational Mode Decomposition for Power Line Interference Removal in ECG Signals

Neethu Mohan, Sachin Kumar S, Prabaharan Poornachandran, Soman K.P

Abstract


Power line interferences (PLI) occurring at 50/60 Hz can corrupt the biomedical recordings like ECG signals and which leads to an improper diagnosis of disease conditions. Proper interference cancellation techniques are therefore required for the removal of these power line disturbances from biomedical recordings. The non-linear time varying characteristics of biomedical signals make the interference removal a difficult task without compromising the actual signal characteristics. In this paper, a modified variational mode decomposition based approach is proposed for PLI removal from the ECG signals. In this approach, the central frequency of an intrinsic mode function is fixed corresponding to the normalized power line disturbance frequency. The experimental results show that the PLI interference is exactly captured both in magnitude and phase and are removed. The proposed approach is experimented with ECG signal records from MIT-BIH Arrhythmia database and compared with traditional notch filtering.

Keywords


Signal processing; power line interference removal; ECG signals

Full Text:

PDF


DOI: http://doi.org/10.11591/ijece.v6i1.pp151-159